Guidance and manufacturer’s declaration - Electromagnetic Immunity 
The 
  Mobile C-Arm X-Ray Product is  intended to 
electromagnetic environment. The purchaser or user of the Brivo OEC 715/785/865 Mobile C-Arm X-Ray 
Product should assure that it is used in an electromagnetic environment as described below: 
Test Level 
Electromagnetic Environment  
IEC 61000-4-6 
 
 
 
 
 
 
Radiated RF 
IEC 61000-4-3 
 
 
 
 
 
 
 
 
at ISM band 
150GHz to 
80MHz 
 
 
 
 
3V /m 80MHz to 
2.5GHz (Ed3.0) 
3V/m 80MHz to 
2.7GHz, 28V/m 
(Ed4.0) 
 
 
 
ISM band 150GHz 
to 80MHz 
 
 
 
 
 
3V/m 80MHz to 
2.5GHz (Ed3.0) 
3V/m 80MHz to 
2.7GHz, 28V/m 
(Ed4.0) 
 
 
 
 
Portable and mobile RF communications 
equipment are used no closer to any part of 
the 
Brivo OEC 715/785/865 Mobile C-Arm X-
Ray Product, including cables, than the 
recommended separation distance calculated 
from the equation appropriate for the 
frequency of the transmitter.  
Recommended separation distance 
 150 kHz to 80 MHz 
    80 MHz to 800 MHz 
       800 MHz to 2.5 GHz 
The formulas are applicable for Ed3.0 only. 
Note: P is the power rating of the transmitter 
in watts (W) according to the transmitter 
manufacturer and d is the recommended 
separation distance in meters (m). 
Field strengths from fixed RF transmitters, as 
determined by an electromagnetic site survey
 
a
, are less than the compliance level in each 
frequency range
 b
.
 
 
Interference may occur in the vicinity of 
equipment marked with the following symbol: 
 
a: Field strengths from fixed transmitters, such as base stations for cellular telephones and land mobile radios, amateur 
radio, AM and FM radio broadcast, and TV broadcast cannot be estimated accurately. To assess the electromagnetic 
environment due to fixed RF transmitters, an electromagnetic site survey should be performed.  If the measured field 
strength exceeds the RF compliance level above, observe the Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product to 
verify normal operation in each use location. If abnormal performance is observed, additional measures may be 
necessary, such as re-orienting or relocating the [EQUIPMENT and/or SYSTEM]. 
b: Over the frequency range 150 kHz to 80 MHz, field strengths are less than 3 V/m. 
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. 
NOTE 2:  These  guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and 
reflection from structures, objects and people. 
Note3: These are guidelines. Actual conditions may vary.