Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
36
Restore Power after Service
1. Re-inspect work area for hazards.
2. Reinstall Workstation covers.
3. Turn on breaker on the Workstation.
4. Connect main interconnect cable to the C-arm.
5. Notify all affected personnel that power is being restored.
6. Remove red lock and tag.
7. Energize the system.
8. Verify proper system operation.
9. Inform customer that system is ready to be returned to service.
High voltage cables, electrolytic capacitors, can retain a dangerous static charge for long
periods after power has been removed. Some devices can acquire a charge spontaneously
without direct contact with other circuitry. Do not touch these components unless power has
been completely removed and they have been discharged. Equipment that contains high
power electrical components must be serviced only by personnel familiar with proper safety
procedures for working near these components. Disconnect AC power before discharging
static electricity in components such as electrolytic capacitors and high voltage cables. Failure
to heed this message will result in death or severe personal injury. Allow adequate time for
static charges to discharge through bleeder resistors. Use a high wattage resistor when
discharging circuits to avoid burns. Have someone watch while you work near high voltages.
This person must remain clear of all circuitry and be prepared to turn off the power to the
system and render aid in an emergency.