Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-18
MCB functional Block
MCU function
- Read the registers of peripheral controller to get the status of sub-systems, such as getting the alarm
status, etc.
- Write the registers of peripheral controller to control the status of sub-systems, such as PIO
(Programmable Input Output), light indicator, buzzer, etc.
- Sample watchdog bite signal from supervisor circuit to check peripheral is still working
- Check all switch signals to respond user’s input
- Communicate with Control Panel board to get the key input and control the LED output
- A/D convert all analog signal, such as shutter location, temperature of oil tank, mA feedback, KV
feedback, etc.
- Communicate with DAC to control mA and KV output
- Communication with Workstation
- Communication with peripheral controller to report the working status of MCU
Peripheral controller function
- Sample the status of input pins, and record them into status registers polled by MCU
- Set output pins according to the control registers accessed by MCU
- Generate interrupt signal of alarm to MCU in order to request the processing
- Generate exposure control signals
- Check switch stuck error
- Control mAS control signals
Operational amplifier circuit function
- Amplify or adjust input analog signals, such as kV feedback, mA feedback, temperature, shutter
location sample signal, and output those signals to the ADC unit (Analog Digital Converter)
- Amplify or adjust analog signals output from DAC (Digital Analog Converter), such as kV setting, mA
setting, mA preheat, and output those signals to other sub-systems
- mA PI control
Transceiver
- Transceiver circuit is responsible for regulating voltage level and supplying driving current.