Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-44
Door_Open_Active_A
Input to show the status of door status
1
Analog
0VAC/24VAC
Input
Door_Open_Active_B
11
Analog
PDU Interface
Signals connected to PDU are listed as follows.
Name
Description
Pin
Attribution
Scale
Comment
Sys_Pwr_Req
Power off requeset
1
Digital
0V/12V
High=effective
Input
Mains_Lost_Err
Status of Mainframe Power
3
Digital
0V/12V
High=effective
Input
Sec_Ovr_Vltge_Err
Transformer secondary over voltage
4
Digital
0V/12V
High=effective
Input
Sec_Udr_Vltge_Err
Transformer secondary under voltage
5
Digital
0V/12V
High=effective
Input
12V_Err
Computer +12V error
6
Digital
0V/12V
High=effective
Input
Rdy_Err
Computer ready error
7
Digital
0V/12V
High=effective
Input
MF_Pwr
Information to indicate MF power status
8
Digital
0V/12V
High=effective
Input
PDU_Connect_2
Information to indicate cable between SIB and
PDU plug or not.
9
Digital
0V/12V
High=effective
Input
PDU_Connect_1
Information to indicate cable between SIB and
PDU plug or not.
10
Digital
0V/12V
High=effective
Input
Beep_Invalid_Key_Cm
d
Invalid key beep command
19
Digital
0V/12V
High=effective
Output
Intlc_Board
Interlock status of EE boards loop
20
Digital
0V/12V
High=effective
Input
ACK_PDU2SIB
ACK of commands send from SIB
21
Digital
0V/12V
High=effective
Input
Rsrv_From_PDU_1
Reserved IOs
11
Digital
0V/12V
High=effective
Input
Rsrv_From_PDU_2
Reserved IOs
12
Digital
0V/12V
High=effective
Input
Rsrv_From_PDU_3
Reserved IOs
13
Digital
0V/12V
High=effective
Input
GND
GND
14
GND
0V
GND
GND
GND
2
GND
0V
GND
Sys_Pwr_Off_Cmd
Power off command from computer
22
Digital
0V/12V
High=effective
Output
Sys_Pwr_Off_Deny
Power off deny from computer
23
Digital
0V/12V
High=effective
Output
MF_Pwr_On_Cmd
MF power on command from computer
24
Digital
0V/12V
High=effective
Output
Rdy
Computer ready
25
Digital
0V/12V
High=effective
Output
Beep_Err_Cmd
Error beep command
26
Digital
0V/12V
High=effective
Output
Rsrv_From_SIB_1
Reserved IOs
15
Digital
0V/12V
High=effective
Output
Rsrv_From_SIB_2
Reserved IOs
16
Digital
0V/12V
High=effective
Output
Rsrv_From_SIB_3
Reserved IOs
17
Digital
0V/12V
High=effective
Output

Table of Contents