7-12 Section 7-3 - Performing Front End (FE) Diagnostics
5) Review the test results in the Comment area and proceed as follows:
If no problems were identified, proceed to re-boot the system, as prompted (Figure 7-6, above).
If no problems were identified but the problem persists, contact your local GE Medical Systems
field engineer for assistance.
If errors were identified but the results were marginal, perform calibration, as described in the
Calibration section, on page 6-11, and then repeat the RFI test.
If errors were identified and the results were not marginal, verify that the RFI cables are
connected and functioning properly, as described in Chapter 5 - Components and Function
(Theory). If the problem persists, contact your local GE Medical Systems field engineer for
assistance.
Figure 7-6 RFI Tab - Self Test Complete