GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
7-126 Section 7-5 - Common Service Interface
iLinq IP Restore Procedure:
When required (for example, after software re-installation), you can perform Restore of iLinq IP settings
from the hard disk, as follows:
1) From the System Configuration window (under the Technical Support Tab - refer to Figure 7-99,
above) trackball to iLinq Settings: Restore and press Select.
A blank screen appears; wait for a while until the phone and modem options message is displayed
then T
RACKBALL to OK and press Select.
2) Wait while the restore process is running - this may take a while.
When complete, a message is displayed prompting for system shutdown.
3.) Perform a full system shutdown, then re-start the Vivid™ 4 system.
The restored iLinq IP settings will now be operational.