Chapter 8 - Replacement Procedures  8-91
13) Using a combination wrench, release the gas spring cable from the cable retaining bracket by 
loosening the securing nut, as shown in Figure 8-74 below..
14.)Release the gas spring cable from the up/down/swivel handle by carefully sliding the cable end 
stopper out through the grooved slot on the handle assembly (refer to the detail shown in 
Figure 8-88 on page 8-101).
NOTE: IMPORTANT - Before finally removing the gas spring cable, pay attention to its path inside the 
machine, and the position of each of its parts.
15.)Remove the gas spring cable.
Figure 8-74 Releasing Gas Spring Cable from Retaining Bracket
Gas Spring Cable
Release securing nut here
Cable retaining bracket