IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-96 Section 8-6 - Lower Section Components Replacement
10) Using a combination wrench, release the swivel release cable from the cable retaining bracket by
loosening the securing nut, as shown in Figure 8-74 below.
11.)Release the swivel release cable from the up/down handle by carefully sliding the cable end stopper
out through the grooved slot on the handle assembly (refer to the detail shown in Figure 8-88 on
page 8-101).
NOTE: IMPORTANT - Before finally removing the swivel release cable, pay attention to its path inside
the machine, and the position of each of its parts.
12.)Remove the swivel release cable.
Figure 8-79 Swivel Release Cable and Retaining Bracket
Figure 8-80 Releasing the Swivel Release Cable Securing Nut
Slide bracket in this direction
Loosen cap screws here
Swivel release cable
Cable retaining bracket
Swivel Release Cable
Cable retaining bracket