IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-104 Section 8-6 - Lower Section Components Replacement
9) Remove the retaining washer [K] from the pin [L] at the lower end of the gas spring.
10) Insert the Hex M8x80mm screw in the appropriate hole to the left of the compartment at its base, 
and then tighten the screw until the pin [L] at the lower end of the gas spring is loose enough to be 
removed, and remove the pin.
NOTE: For clarification, the letters appearing in square brackets [ ] in the following step correspond to the letters 
shown in Figure 8-92 below.
11) Remove the retaining washer [M] from the pin [N] at the upper end of the gas spring, and then 
remove the pin.
12) Carefully remove the gas spring.
Figure 8-92 Gas Spring Lower End Cable - Close-up View
Figure 8-93 Gas Spring - Upper End