Chapter 8 - Replacement Procedures 8-107
.
Figure 8-94 Releasing the Spring Washer from the Center of the Transverse Arm
3) Remove the four bolts (three wrench bolts and one Allen bolt) that secure the front wheel to the
chassis (see Figure 8-94 below) and remove the wheel.
Figure 8-95 Front Wheel (Part No 2269437) Bolted to Chassis