GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
8-116 Section 8-8 - Peripherals
2.) Connect the Remote cable, the B/W Video IN cable, and the Power cable, as shown in
Figure 8-103 below.
3.) Place the B/W printer in its correct final position.
4.) Secure the B/W printer by fastening the belt around it, or together with the VCR (if present).
5.) If more secure fastening is desired, install the optional shelf (P/N H45011CN), which can be ordered
separately if the Color Printer option is not installed.
6.) Re-install the Connectors Panels Cover as described in the Connector Panels Cover Installation
Procedure on page 8 - 8.
7) Reconnect the mains AC power cable from the wall outlet.
8) Turn ON the main circuit breaker located on rear of the Vivid™ 4.
9) Turn ON the Vivid™ 4 by pressing the ON/OFF button on the Control Console.
10.)Select the activation key under System Configuration, VCR/ECG Tab, A/B Swap.
By default, Print B key will be B&W Video printer and Print A key will be Color Video printer. For
more information on switching printer settings, see VCR/ECG Tab on page 3 - 45.
Figure 8-103 B/W Video Printer Connection Cables
Remote cable:
From: Left Connectors Panel,
Print Trig
To: REMOTE
Video IN Cable:
From: Left Connectors Panel,
B/W Video OUT
To: VIDEO IN
Power cable
From: Right Connectors Panel, AC outlet
To: AC IN