GE MEDICAL SYSTEMS
D
IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
Chapter 3 - Installation 3-7
12) Remove the two monitor foam supports (item 12 in Figure 3-4), located one on each side of the monitor.
13) Remove the three silica gel (moisture absorbing) bags (item 5 in Figure 3-4, above) located under
the unit.
14) Raise the control console (monitor) by pressing the release grip located under the unit’s front handle.
15) Remove the two cartons containing the external cables and accessories located under the control
console, and then lower the control console.
16) Lay the top cover (removed from the crate in step 8 on page 3-6) on the floor and push it over the
narrow side of the base, in front of the unit, to act as a ramp (item 3 in Figure 3-5 below).
Figure 3-4 Antistatic Cover, Supports and Accessories
Figure 3-5 Removing the Unit from the Wooden Crate