IRECTION 2392751-100, REVISION 3VIVID™ 4 SERVICE MANUAL
3-12 Section 3-3 - Receiving and Unpacking the Equipment
9) Release the front stopper (item C in Figure 3-10 below) by sliding the metal sleeve located at the
locking pin’s base upwards, and then turning it towards you.
10) Remove the three silica gel (moisture absorbing) bags (item 7 in Figure 3-10, above) located
underneath the ultrasound unit.
11) Raise the control console (monitor) by pressing the release grip located under the unit’s front handle
and remove the two cartons (item 8 in Figure 3-8 on page 3-11) from underneath the control
console. These cartons contain the external cables and accessories.
12) Lower the control console to its lowest position.
13) Remove the manuals (item 4 in Figure 3-8 on page 3-11) from the compartment on the left side of
the ultrasound unit.
14) Press the brake pedal, located near the bottom of the front of the unit, to the right, to prevent the
wheels from swiveling, as described in the Swivel Brake Label section, page 1-23.
15) Carefully pull the unit backwards from the platform base onto the ramp and roll it down to the floor.
16) Press the brake pedal, located near the bottom of the front of the unit, to the left, to lock the wheels,
as described in the Parking Label section, page 1-23.
17) Verify the contents of the shipping carton, as described in Verifying the Shipping Crate Contents on
page 3 - 13.
Figure 3-10 Front Stopper Location