9.3 Communication Methods
137
99
*1: Only the MAV bit (bit 4) is cleared.
*2: All bits except the MAV bit are cleared.
*3: Except the PON bit (bit 7).
During communications, “RMT is lit to indicate the remote control state.
To cancel the Remote state
9.3.5 Initialization Items
: initialized/ ×: not initialized
Initialization method
Item
Power on
RST
command
Device
clear
CLS
command
Device specific functions
(ranges etc.)
× ××
Output queue × ×
Input buffer × ×
Status byte register
×
×
*1
*2
Event register
*3
××
Enable register ×××
Headers on/off ●● ××
9.3.6 Local Function
The “RMT” LED goes out.
+
Remote state cannot be cancelled when the GP-IB command Local Lock Out is
set.