4.4 Displaying the Test Result
63
4
Test is unsuccessful when lights up. display will be retained.
(default condition)
The instrument switches to the FAIL state and stops outputting a voltage if the
measured resistance value deviates from the upper or lower-limit test value.
When the FAIL status is retained, to return to the READY state, press the
STOP key, which will cancel the FAIL Hold function. In normal tests,
the time
from when the instrument achieves and detects a FAIL evaluation to the time
when the output voltage becomes OFF depends on the size of the capacity
property of the tested object.
4.4.2 "FAIL" Determination
To delete the held value shown when the test ended in FAIL.
See "6.2.2 FAIL Hold" (p. 86)
See "6.7 Insulation-Resistance Test Termination Mode" (p. 92)
Voltage-value in the FAIL state
Indicates resistance measurement value
higher than the upper-limit test value
Indicates resistance measurement value
lower than the lower-limit test value
Indicates test abnormalities such as the
output voltage not reaching the set volt-
age.
Resistance-value in the FAIL state