6.2 Display Hold (Hold Function)
85
6
See "Appendix2 List of Optional Functions" (p. A5)
Retains the value shown when the test has ended and PASS is displayed. To
return to the READY state, press the STOP key, which will cancel the Hold.
When the PASS hold function is disabled, the test result is displayed for 0.3 sec-
onds, and the instrument reverts to the READY state.
6.2.1 PASS Hold
1.
( is lit)
Make sure the lamp is lit on the screen.
* will not be displayed during the Double Action setting.
2.
Press the STOP key while holding down the SHIFT key to display the
optional screen (page 1).
3.
Press the / keys to select the setting PASS hold.
0: PASS hold is not used (Initial setting)
1: PASS hold is used
4.
Press the STOP key while holding down the SHIFT key to complete the
setting value.
The instrument reverts to the READY state.
To continue to the setting for the next item, press the key.
SHIFT
STOP
+
or
<Ex.> Select PASS hold.
SHIFT
STOP
+