6.4 Continue testing even after FAIL result is shown (Continuous Test Mode, Only Withstanding-
89
6
See "Appendix2 List of Optional Functions" (p. A5)
Set the instrument to continue or discontinue the test when a NG test result is
shown.
Under this function, the test will continue with lighting up next to UPPER
or LOWER until the test time has run out, except for under the following condi-
tions. When the test is continued under an NG result, the instrument will buzz
and stop the test under the following conditions.
(1) When a current higher than about 24 mA (20 mA-range) or about 12 mA (10
mA-range) is detected in a withstanding test (UPPER FAIL).
(2) When broken wire or poor connection is detected in Contact Check function
(Err message)
See"3.1.7" "4.1.5 Setting the Confirmation Voltage used for upper and lower-limits for
Contact Check"
When FAIL hold is set, the test will hold when the test time has run out.
6.4 Continue testing even after FAIL result is
shown
(Continuous Test Mode, Only Withstanding-Test)
1.
( is lit)
Make sure the lamp is lit on the screen.
* will not be displayed during the Double Action setting.
2.
Press the STOP key while holding down the SHIFT key to display the
optional screen (page 1).
3.
Press the key to move the flashing cursor to the position of page, and
then press the
/ keys to display the optional screen (page 2).
4.
Press the key to move the flashing cursor to the position of Continu-
ous Test Mode.
5.
Press the / keys to select the mode.
0: Continuous Test Mode OFF
(Test completes when an NG result is shown) (Initial setting)
1: Continuous Test Mode ON
(
Continue testing even when NG result is shown)
SHIFT
STOP
+
or
<Ex.> Select Continuous Test ON mode.