11.4 System Reset
198
Contents to be reset
When the system is reset, the following will be initialized. Saved setting data will
also be initialized. Optional settings will restore to the factory default values (ini-
tial settings).
The "
:SYStem:RESet" command starts the same action.
Initial values are as follows.
*items are items initialized with "∗RST" command.
11.4 System Reset
Contents to be reset Initial Values
Test mode
Test mode
Withstand-volt-
age mode *
Withstand-voltage mode
test-voltage 0.20 kV *
upper-limit test value 0.2 mA *
lower-limit test value ON/OFF OFF *
lower-limit test value 0.1 mA *
test-time ON/OFF ON *
test-time 0.3 s *
ramp-up timer ON/OFF OFF *
ramp-up timer 0.1 s *
ramp-down time ON/OFF OFF *
ramp-down time 0.1 s *
test-frequency 50 Hz *
initial voltage for ramp-up 0.0 *
upper-limit voltage for Contact Check ON/OFF OFF *
upper-limit voltage for Contact Check 0.30 kV *
lower-limit voltage for Contact Check ON/OFF OFF *
lower-limit voltage for Contact Check 0.20 kV *
Insulation resistance mode
test-voltage 500 V *
lower-limit test value 1000 MΩ *
upper-limit test value ON/OFF OFF *
upper-limit test value 2000 MΩ *
test-time ON/OFF ON *
test-time 0.3 s *
delay time ON/OFF OFF *
delay time 0.1 s *
upper-limit voltage for Contact Check ON/OFF OFF *
upper-limit voltage for Contact Check 600 V *
lower-limit voltage for Contact Check ON/OFF OFF *
lower-limit voltage for Contact Check 500 V *