11.4 System Reset
199
10
11
Procedure for System Reset
Turn ON the main power switch while pressing the SHIFT key.
Press and hold the SHIFT key until is shown.
Optional Functions
FAIL Hold, Double Action,
START protection function
1*
Output-voltage restricting value 5.0 kV *
PC interface 0
GP-IB address 3
Inter-lock 0
Interface command "START" 0
Others 0 *
Status Out
All 0 (Not set) *
Saved test conditions
Files in the withstand-voltage mode and insulation-resistance mode
will be initialized respectively.
Saved data
Results of the withstand-voltage test and the insulation-resistance
test will be initialized.
Contents to be reset Initial Values