EasyManua.ls Logo

Hioki 3174 - 3.4 Displaying the Test Result

Hioki 3174
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.1 Setting the Test-Conditions
38
Confirmation Voltage used for upper and lower-limits for Contact Check.
Contact Check can be carried out by measuring the terminal voltage of the
tested object. When this function is pre-set, test abnormalities can be detected
whenever the output voltage deviates from the pre-set ranges. The setting for
the upper or lower limit voltage confirmation is the same as that used for the H
side terminal Contact Check, but setting either the upper limit or the lower limit
will also automatically allow for Contact Checks on the L side terminals.
By measuring the voltage on the H side terminals of the tested object, checks on
whether an accurate voltage is supplied to the test terminals or not (whether the
test lead is properly connected or not: Contact Check) can be carried out. When
a value outside the range of the pre-set terminal voltage is achieved, an error
number will show and the test will terminate.
When Contact Check Function is not in use, press the ON/OFF key to set to
"OFF".
3.1.7 Setting the Confirmation Voltage used for upper
and lower-limits for Contact Check
1.
Press the key to move the flashing cursor to the upper-limit confirma-
tion voltage for Contact Check.
2.
Press the / keys to set the upper-limit confirmation voltage for Con-
tact Check.
Setting range: 0.20 to 5.00 kV (The value changes by 0.01 kV.)
To change the value by 0.1 kV, while holding down the SHIFT key, press /
keys.
If no upper-limit confirmation voltage for Contact Check is required, set it to
“OFF” using the ON/OFF key.
3.
Press the key to move the flashing cursor to the lower-limit confirma-
tion voltage for Contact Check.
or
<Ex.> Set to 0.4 kV
.

Table of Contents

Related product manuals