6.9 Status Out
96
See "Appendix2 List of Optional Functions" (p. A5)
The status out relay can be switched ON or OFF by setting a few conditions. The
following are the conditions that can be set.
The status out relay connection will turn ON as long as any one of the conditions
is met.
H.V.ON High voltage output. Shown together with DANGER lamp.
TEST Testing. When lights up or flashes.
PASS PASS status. When lights up.
FAIL FAIL status. When lights up.
INT.LOCK Interlock status.
READY READY status. When lights up.
EXT.CONT. When controlled externally. Controlled by external I/O and remote
control box (“EXT” lights up on the screen).
POWER.ON When the power is switched on.
All conditions are invalid in the initial setting.
6.9 Status Out
Do not connect any device which requires a current of more than 0.5 A to the
relay terminal for status out. It may cause the device to malfunction.
Specifications for Relay Terminal
Maximum input
voltage
150 V AC, 30 V DC
Maximum contact
current
0.5 A
Rated wire Single wire : φ1.2 mm (AWG16)
Stranded wire: 1.25 mm
2
(AWG16),
wire diameter 0.18 mm or more
Usable wire Single wire : φ0.4 to φ1.2 mm (AWG26 to 16)
Stranded wire: 0.25 to 1.25 mm
2
(AWG22 to 16)
Wire diameter 0.18 mm or more
Standard length of
removed wire sec-
tion
11 mm