4.4 Displaying the Test Result
62
Test is successful when lights up. Under the settings in the insulation-
resistance test termination mode, when the resistance measurement value is
higher than the lower-limit test value, and lower than the upper-limit test
value, a
PASS will show and the voltage output will stop.
In the initial setting, will light up for about 0.3 seconds before returning to
READY mode.
Installation Flow of PASS determination
(Timer stops: See "6.7 Insulation-Resistance Test Termination
Mode" (p. 92))
4.4.1 "PASS" Determination
To hold the display
See "6.2.1 PASS Hold" (p. 85)
See "6.7 Insulation-Resistance Test Termination Mode" (p. 92)
Voltage-value in the PASS state Resistance-value in the PASS state
READY TEST
PASS
DANGER is lit
START
Test voltage
Set delay time
(Undetermined
period)
Set test time
Discharge
time
60 V DC
Time
TEST
flashes
Time to
increase
the voltage