6.2 Display Hold (Hold Function)
87
6
See "Appendix2 List of Optional Functions" (p. A5)
Retains the value when the test has been forcibly terminated by pressing the
STOP key during the test. To return to the READY state, press the STOP key,
which will cancel the Hold.
If the Hold function is not selected, the instrument switches to the READY state
upon forced termination of the test.
When the "Evaluate even for forced termination of test" is set for the optional
function, this setting will be ignored.
6.2.3 STOP Hold
1.
( is lit)
Make sure the lamp is lit on the screen.
* will not be displayed during the Double Action setting.
2.
Press the STOP key while holding down the SHIFT key to display the
optional screen (page 1).
3.
Press the key to move the flashing cursor to the position of STOP hold.
4.
Press the / keys to select the mode.
0: STOP hold is not used (Initial setting)
1: STOP hold is used
5.
Press the STOP key while holding down the SHIFT key to complete the
setting value.
The instrument reverts to the READY state.
To continue to the setting for the next item, press the key.
SHIFT
STOP
+
or
<Ex.> Select STOP hold.
SHIFT
STOP
+