EasyManua.ls Logo

Hioki 3174 - Error Indication

Hioki 3174
224 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
11.3 Error Indication
197
10
11
If an error occurs, the 3174 displays the following on the screen.
If the displayed number is outside the test setting range, follow the instructions
for "
Err3" in the table above.
When errors other than the above are displayed, the instrument may be mal-
functioning. Contact your dealer or Hioki representative.
11.3 Error Indication
Message Description Remedy
Err0 Interlock state.
See "8.3 Inter-lock Function" (p. 111)
Err1
External switch is
unusual.
Other switches apart from the Remote
Control Box (9613, 9614) are connected.
If message does not go off even when
standard Remote Control Box is con-
nected, damage is possible. Contact your
dealer or Hioki representative.
Err2
Damage due to EXT
I/O is possible.
Contact your dealer or Hioki representa-
tive.
Err3
Damage due to
DANGER lamp is
possible.
Contact your dealer or Hioki representa-
tive.
Err4
Contact error at the L
side terminal.
Confirm the connection to the instru-
ment
.
Err5
Test voltage is set
lower than lower-limit
voltage.
Confirm the setting.
Err6
Output measure-
ment voltage is lower
than the lower-limit
voltage.
An output problem is possible. Check the
output voltage before contacting your
dealer or Hioki representative.
Err7
Contact voltage is
lower than the lower-
limit voltage.
A measurement-type problem or a con-
tact error is possible. If the connection on
the high voltage side does not present
any problem, contact your dealer or Hioki
representative.
Err8
Set voltage is set
higher than the
upper-limit voltage.
Confirm the setting.
Err9
Output voltage is
higher than the
upper-limit voltage.
An output-type or measurement-type
problem is possible. Contact your dealer
or Hioki representative.
ErrH
An internal part of the
instrument is over-
heated.
Leave the instrument for several minutes
without turning OFF the power switch.
(Cool the instrument.) If the instrument is
not restored, it may be malfunctioning.
Contact your dealer or Hioki representa-
tive.

Table of Contents