Section 1 - Introduction
1-9
SPECIFICATION MODEL DEFINITION
Operating
Temperature:
All 15
o
- 32
o
C (60
o
- 90
o
F)
Humidity: All 20 - 80% relative Humidity, non-condensing
Footprint: Length Width Height
m inches m inches m inches
C-arm not
rotated, table not
extended
A 2.02 79.5 1.40 55 1.42 56
C-arm rotated,
table extended
A 3.02 119 1.50 59 1.42 56
C-arm not rotated
SL 2.02 79.5 1.40 55 1.42 56
C-arm rotated
SL 2.02 79.5 1.50 59 1.42 56
table extended
W 3.02 119 1.50 59 1.42 56
table not
extended
W 2.02 79.5 1.22 48 1.42 56
C 2.02 79.5 1.40 55 1.42 56
Average Heat
Load:
ALL 1000w (3400 BTU/hr)
A,SL Adjustable, 71cm (28") from floor when scanning in AP mode
Patient Table
Height:
W,C 71cm (28”)
Positioning Laser: All
Laser Diode (<1mW) cross hair, with emergency mechanical
shutter
A,SL Selectable by scan type
X-ray Collimation:
W,C 1.0mm slit
A,SL
Exam table is capable of moving +/-2.54cm (1.0") in the x-axis
from center location with scan arm in lateral position.
Lateral Tracking:
C,W N/A
Leakage Current: All Normal <75µA Single Fault <400µA
A/SL 1 line pair/mm (approximately 0.5mm)
Resolution:
C/W 0.5 line pair/mm (approximately 1.0mm)