Section 3 - Installation
3-29
THE RADIATION MEASUREMENT REPORT
After installation, the field engineer must fill out the Radiation Measurement Report and keep
this information on file. A sample of this report follows:
QDR 4500 Radiation Measurement Report
F/E Signature Date
Voltage and Current Tests
Radiation Tests
LI
Customer Name
Customer Address
Customer Contact
Hologic Model Number
Serial Number
Date Measurement Taken
Line Voltage:
(choose one)
Tube Peak Potential:
Tube Current: Reproducibility Test:
SD Hi Air
(14)
!
SD LoAir
(14)
!
Leakage Test
using Victoreen Model #450P (or equivalent) S/N
Dosage Test
using Victoreen Model #450P (or equivalent) S/N
Array Beam
mR
Array Beam
mR/h
Array Beam
mR/h
(100V) ( 90V, 110V)
" !
(120V) ( 108V, 132V)
" !
(230V) (207V, )
" !
253V
(80 kVp)
( , 88kVp)
"
72 kVp
!
(100 kVp) (90kVp, )
" !
110kVp
(140 kVp)
( 126kVp, )
" !
154kVp
Mode 1 (3 mA)
( 6.5mA, )
" !
13.5mAMode 3 (10 mA)
( 2.0mA, )
" !
4.0mA
Scatter Test
using Victoreen Model #450P (or equivalent) S/N
CSD-0026-01 Rev B