QDR
4500 Technical Manual
6-44
or
Unable to Read File: ________
or
Unable to Read Line by Line d0's
or
Unable to Read Previous Analysis Results
or
Unable to Rewind Q-File
or
Unable to Rewind File: ________
or
Unable to Read BMD Image Frame
or
Unable to Read HALO Frame Header
or
Unable to Read Q-Data File
or
Unable to Write to Q-Image File
Hardware or
Software Error:
These messages indicate problems performing I/O operations (usually to the hard drive,
but possibly to diskette or optical).
Action: Ensure that there is adequate space on the hard drive (Check the "Room for
..." message in the status window. You should always leave room for 1 or 2
scans). There may be a disk problem; appropriate diagnostics should be
run. If you are unable to find a hardware problem, the message should be
reported to Hologic Technical Support. Include as much information as
possible about what you were doing at the time.