QDR
4500 Technical Manual
6-46
Unknown ROI type in DrawROI
Software Error: These messages should be reported to Hologic Technical Support. Include
as much information as possible about what you were doing at the time.
device:
Unrecognized Command Code
Software Error: This message should be reported to Hologic Technical Support. Include as
much information as possible about what you were doing at the time.
Unrecognized device command [E_UNKNOWN_COMMAND_I]
Comment: Returned by all devices in response to an unrecognized command.
Action: Report this error to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
Unsolicited message from C/C [E_UNSOLICITED_MESSAGE_I]
Action: Report this error to Hologic Technical Support. Include as much
information as possible about what you were doing at the time.
X-ray controller is not generating A/C line interrupts
[E_NO_TIMER_INTERRUPT_I]
LOCUS: DAS A request was made to read the a/d's and there was no system line frequency
interrupts (required to synchronize DAS with C-arm).
LOCUS: Global The system was on-line and another on-line command was issued but there
were still no timer interrupts.
Action: Check that the Instrument Power switch and the X-Ray Enable key are on.
Also, see the discussion of E_INTERLOCK_INHIBIT_I,
E_EMERGENCY_STOP_I and E_COLLISION_IMMINENT_I.
Refer to the
No A/C Line Interrupts
header, page 6-17 in this section.