The Injection Counter/Limit are used to monitor the GC and predict when the GC septa or
injection liner should be replaced.
Please note:
The counters are automatically reset when the warning message is cleared;
The Plunger Stroke Counter is automatically reset after a new syringe installation.
If SyringeID system is enabled (Liquid mode), Plunger Strokes Counter reset is not
allowed (see paragraph 4.2.6.8 “Setup configuration: SyringeID”).
Select Exit to exit from this screen and return to the “Setup Configuration”” screen
(see paragraph 4.2.6 “Setup: Configuration”):
4.2.6.6 Setup configuration: Advanced
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>NEXT>ADVANCED
This menu is password protected and is only accessible to service representatives.
4.2.6.7 Setup configuration: Activation
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>NEXT>Activation
This screen contains a list of 9 optional features (from 1 to 9). Select a feature to enable it (to
know if there feature available for activation, contact your sales representative).
A screen showing the following message: You are activating the feature X(“X” is the feature
number) appears.
Tap CANCEL to return to the previous screen.
Tap TEMPORARY to temporarily enable the selected feature (see below how to
proceed).
Tap FULL, to definitively enable the selected feature. A password is required for the
activation. Please contact your sales representative to get the password.
If you have tapped “TEMPORARY a screen showing the following message: “Random
number: X” appears. If this is your first activation attempt, send this random number to your
sales representative to obtain the password for the temporary activation. Then tap “CONTINUE”
to access the screen for the password entry and temporarily enable the feature. If you haven't
received the password yet, from the screen for the password entry you can tap “CANCELand
retry the activation later. You have only to re-execute the described operations. The unit will re-
show you the same random number (because you have already tried the activation, but you
haven't complete it). Directly tap ACTIVATEto access the screen for the password entry and
temporarily enable the feature.