HT2800T_en_c.doc Page 134 of 275
Warning
In liquid mode the Bar Code reader option can be enabled only if the proper vial locator
with gripper is mounted (see paragraph 2.1.2.2 “Syringe location for Liquid Mode”).
To see how to mount the vial locator with gripper in liquid mode see paragraph (see
paragraph 3.6.4 “Vial Locator: Installing and removing (Liquid mode)”) .
The following parameters can be viewed/edited:
Enable: enables or disables the Bar Code reading from the vials;
On reading error: sets the autosampler behaviour if the BCR cannot read the Bar Code
of a vial;
Skip
The vial, whose Bar Code has not been read, is not processed
and it is repositioned on the tray. The “Startsignal emission to
the analyzer depends on the “Empty Position Synchrosetting
(4.2.6.3 “Setup configuration: Run“).
Process
The vial is processed, even if the Bar Code has not been read.
Test: starts the Bar Code reading test. This procedure checks the reading of the Bar
Code applied on the indicated vial position. After the test execution, a screen shows the
read Bar Code and the reliability of the reading expressed as a percentage. When the
tray is open, it is possible to verify that the read Bar Code is correct. If the reading fails,
“None” appears instead of the Bar Code. Tap “CONTINUE” to end the test procedure.
Select Exit to exit from this screen and return to the “Setup Configuration”” screen
(see paragraph 4.2.6 “Setup: Configuration”):
4.2.6.10 Setup configuration: Cleaning Device
Available only for HT2850T in SPME mode.
Access sequence: RUN>SETTINGS>SETUP>CONFIGURATION>NEXT>CLEANING DEVICE
This menu is available only for HT2850T. The following parameters can be viewed/edited:
Enable: enables or disables the Fiber Cleaning Device (see paragraph 11 “Appendix A
Glossary“).
Cleaning device Temperature: sets the Fiber Cleaning Device temperature. This value
can be between 210°C and 300°C (active conditioning) or off (inactive conditioning);