Pre Wash Cycle (s): the number of pre washes (washes before injection)
Pre Wash Volume: the quantity (volume) of solvent to use for each pre-washing
operation;
Pre Wash Mode: how often the pre washing occurs (see below);
Post Wash Cycle (s): the number of post washes (washes after injection)
Post Wash Volume: the quantity (volume) of solvent to use for each post-washing
operation;
Post Wash Mode: how often the post washing occurs (see below);
Pre and Post wash mode can have the following values:
Inject
Wash is made every injection
Sample
Wash is made on each new vial: it is not made between
injections of the same sample from the same vial.
Step
Wash is made for each new step of the sequence.
Select Exitto exit from this screen and return to the RUN: METHOD X” screen (see
paragraph 5.2.2 “Method X: Setting method parameters”).