Fiber
lenght
Extraction/
Derivatization
depth
(mm)
0mm
<10 The instrument moves the
needle to have the needle tip just
under the septum, then expose
the fiber moving only the plunger.
Cover
Under septum,
depth = 0
33mm
47mm
0-10mmFiber
10<depth<28 The instrument moves the
needle 28mm over the final
position, then moves the plunger
to expose 10 mm of fiber and if
necessary, up to 18mm of fiber
holder.
Cover
33mm
47mm
0-18mm
F
iber holder
10mm Fiber
10<depth<47 The instrument moves the
needle 28mm over the final
position, then moves the plunger
to expose 10 mm of fiber and
18mm of fiber holder and, if
necessary, it moves the needle
down up to 19mm further.
Un
de
der septum,
pth
= 0
33mm
47mm
0-19mm
Needle
18mm
Fiber holder

Table of Contents