HT2800T_en_c.doc Page 18 of 275
ENVIRONMENT CONDITIONS
Room temperature limits 1C - 35°C
Ambient humidity limits 5% - 80% (not-condensable)
Maximum altitude 2000m
SOUND PRESSURE LEVEL
Maximum measured level 60dBA (below the limits of 85dBA defined by the
regulations in force)
SAFETY INFORMATION
The autosampler is classified as shown below: Pollution degree 2
Overvoltage category II
Devices for use indoors
ANALYTICAL PARAMETERS
HEADSPACE MODE
Tray capacity 42 vials (20ml); optional: 6 and 10ml
Syringe volume 2.5ml (standard); optional:1ml
Maximum pressure in the injector 2.5bar
(2.5ml syringe)
Conditioning
Oven positions 6
Oven Temperature off, from 40°C to 170°C
Incubation time from 0 to 999min step 1min
Shaking method orbital
Shaker speed from very low to very high
Shaking cycles on/off, from 0 to 9.9min
Sampling
Syringe Temperature off, from 40°C to 150°C
Sample volume 0.01ml stepwise
Pull Up Strokes up to 15 strokes
Sample speed from 0.5 to 100ml/min (40ml/min for 1ml syringe)
Injection