HT2800T_en_c.doc Page 183 of 275
Follow this procedure to move from SPME to Liquid mode
Access sequence: RUN>CHANGE MODE>LIQUID
The screen “You are moving to Liquid mode” confirms the start of the changing procedure. Tap
“CONTINUEto start the switching operation or “ABORT to return to the “CHANGE MODE”
screen.
Open the syringe location by pushing up the sliding lid and follow the wizard:
Warning
If syringeID system is used, the syringe information are loaded in a RFID tag included in
the black adapter mounted on the syringe for the syringeID system (see paragraph 11.2
“Liquid mode “). Every adapter must be associated to a syringe only in order to maintain
the info regarding the syringe always correct and updated. Do not remove the syringe
from its black adapter.