HT2800T_en_c.doc Page 206 of 275
Figure 182: Syringe warmer assembly mounting
d) Tap “CONTINUE”.
9) “Select syringe volume” screen: select the volume of the installed syringe and press
“SAVE”.
10)“Remove spacers (and insert vials in position A4, C and System Integrity Tool in proper
position” screen: remove the spacers for 6/10ml vials in the oven (if present), using the
supplied tweezers (see paragraph 2.3 “Tool kit“), an headspace sample vial
13
13
in A.4
position of the rack and in position C of the oven and a the System Integrity Tool (only if
you purchased the GLP pack, see paragraph 1 . 7 . 2 “ Options” ) into its the
dedicated position on the back of the autosampler (see paragraph 2 . 1 . 1
“ HT2800T/HT2850T Overview““), then press “CONTINUE”.
11) If during the final Zero procedure the alignment on injector is not correct please modify it
(see paragraphs 4.2.1.2 “Alignment: Front Injector” and 4.2.1.3 “Alignment: Rear
Injector “).
13
The headspace vial to be put in the rack must have the same volume of the vial used last time in
headspace mode.