EasyManuals Logo

IAI PCON-CB series User Manual

IAI PCON-CB series
538 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #222 background imageLoading...
Page #222 background image
4. Operation
4.2 Operation in Positioner Mode
4-96 ME0342-4B
4.2.8 Preparation for Operation in Force Sensor Pressing
(Loadcell Calibration)
The force sensor pressing is a feature dedicated for PCON-CBP/CGBP.
With the performance of a loadcell for the feedback control of the pressing force during the
pressing operation in operation of the force sensor used pressing (PIO Patterns 6 and 7),
pressing operation with high precision should be available.
Before having the pressing operation of the force sensor used pressing, have the settings and
process described below established.
Caution
For the force sensor used pressing, there is a need of an actuator applicable for the
dedicated loadcell and the force sensor used pressing.
In the calibration, the value at the moment of the calibration should set as the datum
point (0 [N]). Therefore, it is necessary to consider that load of a fixture attached on the
loadcell should only be the load applied on the loadcell during the calibration.
[1] Initial Setting
The pressing operation with the force sensor used pressing requires a use of a loadcell. It is
necessary to have the initial setting established before starting operation. Establish the settings as
shown in the thick-bordered boxes to the four parameters shown in the table below.
No. Name
Setting for
Actuator
Equipped
with
Loadcell
Setting When
Using Force
Sensor Used
Pressing
Remark
92
Loadcell Use Select
[0: Not to Use / 1: To Use]
1 1
0: Loadcell Not to be Used
1: Loadcell to be Used
93
Pressing Control Select
[0: Current Limit / 1: Force Sensor]
1 1
0: Control with Current Limit
(Standard)
1: Control with Force Sensor
117
Automatic Loadcell Calibration at
Startup
[0: Not to Perform / 1: To Perform]
1
1 (Recommended)
or
0
0: Loadcell Calibration Not to
be Performed
Automatically
1: Loadcell Calibration to be
Performed Automatically
118
Pressing Operation When Loadcell
Calibration Incomplete
[0: Prohibit / 1: Permit]
0 0
0: Pressing Operation When
Calibration Incomplete to
be Prohibited
1: Pressing Operation When
Calibration Incomplete to
be Permitted

Table of Contents

Other manuals for IAI PCON-CB series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the IAI PCON-CB series and is the answer not in the manual?

IAI PCON-CB series Specifications

General IconGeneral
BrandIAI
ModelPCON-CB series
CategoryController
LanguageEnglish

Related product manuals