Table 5-4
FORMat command summary
Command Description
Default
parameter
SCPI
:FORMat
[:DATA] <type>[,<length>]
[:DATA]?
:ELEMents <item list>
:ELEMents?
:BORDer <name>
:BORDer?
Select data format: (ASCii, SREal or DREal).
Query data format.
Specify data elements: (READing, CHANnel, and UNITs).
Query data elements.
Select binary byte order: (NORMal or SWAPped).
Query byte order.
ASCii
READing
SWAPped
Table 5-5
ROUTe command summary
Command Description
Default
parameter
SCPI
:ROUTe
:CLOSe <chan num>
:STATe?
:OPEN:ALL
:MULTiple
:CLOSe <list>
:STATe?
:OPEN <list>
:SCAN
[:INTernal] <list>
[:INTernal]?
:EXTernal <list>
:EXTernal?
:LSELect <name>
:LSELect?
Commands to control scanner card:
Close specified channel (1 to 10) or channel pair (1 to 5).
Query closed channel (or channel pair).
Open all input channels (1 through 10).
Path to close and open multiple channels:
Close specified channels (1 to 11).
Query closed channel.
Open specified channels (1 to 11).
Path to scan channels.
Specify internal scan list (2 to 10 channels).
Query internal scan list.
Specify external scan list (2 to 800 channels).
Query external scan list.
Select scan operation (INTernal, EXTernal, or NONE).
Query scan operation.
1-10
1-10
NONE
SCPI Command Reference 5-9

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz