:STATe <b>
[:SENSe[1]]:CURRent:AC:REFerence:STATe <b> Control reference for ACI
[:SENSe[1]]:CURRent[:DC]:REFerence:STATe <b> Control reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence:STATe <b> Control reference for ACV
[:SENSe[1]]:VOLTage[:DC]:REFerence:STATe <b> Control reference for DCV
[:SENSe[1]]:RESistance:REFerence:STATe <b> Control reference for 2
[:SENSe[1]]:FRESistance:REFerence:STATe <b> Control reference for 4
[:SENSe[1]]:FREQuency:REFerence:STATe <b> Control reference for FREQ
[:SENSe[1]]:PERiod:REFerence:STATe <b> Control reference for PER
[:SENSe[1]]:TEMPerature:REFerence:STATe <b> Control reference for TEMP
[:SENSe[1]]:VOLT[:DC]:STERminals:REFerence:STATe <b> Control reference for sense terminals
Parameters <b> = 1 or ON Enable reference
0 or OFF Disable reference
Query :STATe? Query state of reference
Description These commands are used to enable or disable Reference for the specified
function. When enabled, the displayed reading will include the programmed
reference value (see :REFerence <n> and :ACQuire). When disabled, the
displayed reading will not include the reference value.
:ACQuire
[:SENSe[1]]:CURRent:AC:REFerence:ACQuire Acquire reference for ACI
[:SENSe[1]]:CURRent[:DC]:REFerence:ACQuire Acquire reference for DCI
[:SENSe[1]]:VOLTage:AC:REFerence:ACQuire Acquire reference for ACV
[:SENSe[1]]:VOLTage[:DC]:REFerence:ACQuire Acquire reference for DCV
[:SENSe[1]]:RESistance:REFerence:ACQuire Acquire reference for 2
[:SENSe[1]]:FRESistance:REFerence:ACQuire Acquire reference for 4
[:SENSe[1]]:PERiod:REFerence:ACQuire Acquire reference for PER
[:SENSe[1]]:FREQuency:REFerence:ACQuire Acquire reference for FREQ
[:SENSe[1]]:TEMPerature:REFerence:ACQuire Acquire reference for TEMP
[:SENSe[1]]:VOLT:[DC]:STERminals:REFerence:ACQuire Acquire reference for sense terminals
Description When one of these commands is sent, the measured input signal is acquired
and established as the reference value. This command is typically used to
zero the display. For example, if the instrument is displaying a 1µV offset,
sending this command and enabling Reference (see :STATe) zeroes the
display.
This command is functional only if the instrument is on the specified
measurement function. Sending this command while in any other function
causes an error. Also, if the latest reading is overflowed (“OFLO”) or a
reading has not been triggered (“----”), an error occurs when this command
is sent.
The :ACQuire command is coupled to the :REFerence <n> command. See
the description for :REFerence for details.
5-44 SCPI Command Reference

Table of Contents

Other manuals for Keithley 2010

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2010 and is the answer not in the manual?

Keithley 2010 Specifications

General IconGeneral
Frequency Range3 Hz to 300 kHz
DC Voltage Range100 mV to 1000 V
AC Voltage Range100 mV to 750 V
DC Current Range3 A
DC Current Accuracy0.05%
AC Current Range3 A
Resistance Accuracy0.01%
InterfaceGPIB, RS-232
Power Supply50 Hz or 60 Hz