LABORIE Goby Owner’s Manual GOBY-UM02 106
IEC 60601-1-2:2014 Table 1 Requirements
The Goby Urodynamic Analyzer System is intended for use in the electromagnetic environment specified below.
The customer or the user of the Goby Urodynamic Analyzer
System should assure that it is used in such an environment.
Electromagnetic Environment – Guidance
RF Emissions
CISPR 11
Group 1
The Goby Urodynamic Analyzer System uses RF energy only for
its internal function.
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class A
The Goby Urodynamic Analyzer System is suitable for use in all
establishments other than domestic, and may be used in domestic
establishments and those directly connected to the public low-
voltage power supply network that supplies buildings used for
domestic purposes, provided the following warning is heeded:
Warning
: This equipment/system is intended for use by healthcare
professionals only. This equipment/system may cause radio
interference or may disrupt the operation of nearby equipment. It may
be necessary to take mitigation measures, such as re-
orienting or
relocating the Goby Urodynamic Analyzer System or shielding the
location.
Harmonic Emissions
IEC 61000-3-2
Class A
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
Table 14: Table 1 Requirements - Electromagnetic Environment