EasyManua.ls Logo

LABORIE Goby - Calibrate the Infusion Transducer; Calibrate the UPP; Restore Factory Calibration; Figure 114: Pump, Infusion, and UPP Calibration Window - Infusion Section

LABORIE Goby
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LABORIE Goby Owner’s Manual GOBY-UM02 92
7.4 CALIBRATE THE INFUSION TRANSDUCER
1. Plug the infusion transducer cable into the connector at the bottom of the Goby™ Hub.
2. Switch ON the
Goby™ Hub
.
3. Start the UDS120 Goby software and double-click the
Goby™ Hub
icon in the main window to open the Goby
Device Manager window (Figure 108).
4. Click the Setup button in the
Goby™ Hub
section to launch the Setup window. Click the Device Calibration
button (Figure 109).
5. Select Infusion under the Infusion Section option
(Figure 114).
6. Place a full infusion bag on the infusion transducer.
7. Follow the on-
screen instructions to complete the
calibration.
Figure 114: Pump, Infusion, and UPP Calibration
Window Infusion Section
7.5 CALIBRATE THE UPP
1.
Gather a sharpened pencil for marking purposes and a ruler with millimeter (mm) markings.
2. Plug the UPP cable into the connector at the bottom of the
Goby™ Hub
.
3. Switch ON the
Goby™ Hub
.
4. Start the UDS120 Goby software and double-click the
Goby™ Hub
icon in the main window to open the Goby
Device Manager window (Figure 108).
5. Click the Setup button in the
Goby™ Hub
section.
6. In the Setup window click the Device Calibration button (Figure 109).
7. Click UPP under the UPP Section option (Figure 115).
8. Follow the on-screen instructions to
complete the
calibration.
NOTE:
When marking distances on the puller,
mark the
edge of the mobile clamp that stops along the nose tube.
Figure 115: Pump, Infusion, and UPP Calibration
Window UPP Section
7.6 RESTORE FACTORY CALIBRATION
Not available.
Mark edge where
clamp stops along
nose tube

Table of Contents

Related product manuals