EasyManua.ls Logo

LABORIE Goby - Troubleshooting and Appendices; Troubleshooting Common Issues

LABORIE Goby
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LABORIE Goby Owner’s Manual GOBY-UM02 90
7.2 CALIBRATE THE PUMP
1. Switch ON the Goby™ Hub.
2. Start the UDS120 Goby software and double-
click the
Goby™ Hub
icon in the main window
to open the Goby Device Manager window
(
Figure 108).
Figure 108: Access Goby Device Manager Goby™ Hub
3. Click the Setup button to launch the Goby Hub
Setup window. Click the
Device Calibration
button (
Figure 109).
Figure 109: Goby Hub Setup Window
4. Select the Pump option under the Pump Section
(Figure 110).
5. Set up the pump tubing. Refer to the
Set up the
Pump section on page 18.
6. Type in the
Calibration Name
. The name will be
used
when selecting a calibration set before
running a test.
NOTE:
Do not use commas in Calibration Names.
7. Select the Fluid Density and the Catheter Size
that will be used during the Pump calibration
and enter their values in the appropriate fields.
8. Follow the on-screen instructions to complete
the calibration.
Figure 110: Pump, Infusion, and UPP Calibration Window

Table of Contents

Related product manuals