LABORIE Goby Owner’s Manual GOBY-UM02 60
5.6.7 Voiding Summary
The Voiding Summary applies only to tests involving Uroflow information like the Uroflowmetry and pressure-flow
procedures. The Summary compares a patient’s test results with the normal values. Refer to
Table 19: Parameters of
Uroflowmetry 80 Males or Table 20: Parameters of Uroflowmetry 50 Females in the Appendices
section starting on
page 101. Statistical deviations are calculated using these standards and are intended strictly for comparison purposes
and should not be used solely for diagnostic purposes.
IMPORTANT: Since this comparison is based on patient gender, the gender must be correctly entered in the Patient
Information Dialog Box.
To access the Voiding Summary:
Click
Info
from the menu bar and select
Voiding Summary.
The Voiding Summary printout supplies a comparison to the Nomogram
tables (Figure 71). They are provided for both male and female patients
on the same printout.
Figure 71: Voiding Summary Window