EasyManua.ls Logo

LABORIE Goby - 7 Calibration; Calibrate the Urocap IV Uroflowmeter; Figure 105: Access Goby Device Manager - Urocap IV; Figure 106: Urocap Setup - Device Calibration

LABORIE Goby
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
LABORIE Goby Owner’s Manual GOBY-UM02 89
7 CALIBRATION
To calibrate the Urocap™ IV, the UDS Roam, the pump on the Goby™ Hub, the infusion transducer, or the UPP Puller,
open the Goby Device Manager window and select the appropriate device.
IMPORTANT: Plug the Urocap™ IV, the UDS Roam™, and/or the Goby™ Hub into their chargers before beginning
calibration. Make sure the cart is on a level surface and the wheels are locked.
The power switch for the Goby Hub is located at the bottom of the Hub. For an overview of connections for the Urocap
IV and the GobyHub see the Device Connectors and Buttons section on page 30.
7.1 CALIBRATE THE UROCAP IV UROFLOWMETER
1. Plug the Urocap™ IV into the charger and plug the
charger into an outlet.
2. Start the UDS120 Goby software and double-click the
Urocap™ IV
icon in the main window to open the
Goby Device Manager window (
Figure 105).
Figure 105: Access Goby Device Manager Urocap IV
3. Click the Setup button in the Urocap section to launch
the Urocap™ Setup window. Click the
Device
Calibration
button (Figure 106).
Figure 106: Urocap Setup Device Calibration
4. Place an empty beaker on the Urocap™ IV and then
click
Apply
in the UROCAP4 Calibration window
(
Figure 107).
5. Follow the on-
screen instructions to complete the
calibration.
6. Click
Close
once calibration is complete.
Figure 107: UROCAP4 Calibration Window

Table of Contents

Related product manuals