12 Maintenance and Troubleshooting
3. If the system was previously open, manually purge the system before proceeding:
a. Press 8 to open the Expansion valve and 9 to close the Vent valve; X - O - X.
b. Press 7 to open the Fill valve. Once the sample chamber fills to the appropriate pres-
sure, press 7 to close the Fill valve.
c. Press 9 to open the Vent valve; X - O - O.
d. Repeat this procedure two or three times.
e. Press 8 to close the Expansion valve; X - X - O.
4. Press 7 to open the Fill valve; O - X - O.
5. Fill the sample chamber to 19.5 psig.
6. Press 7 to close the Fill valve; X - X - O.
7. Observe the pressure display. After an equilibration period (about 20 to 30 seconds), the pres-
sure should not decrease more than 0.007 psig (0.048 kPag) or increase more than 0.02 psig
(0.14 kPag) in a two minute span.
n If the pressure does not decrease more than 0.007 psig (0.048 kPag) or increase more
than 0.02 psig (0.14 kPag) in a two minute span, proceed to step 8.
n If the pressure decreases more than 0.007 psig (0.048 kPag) or increases more than 0.02
psig (0.14 kPag) in a two minute span, temperature instability or a leak may be indicated.
Vent the system, then repeat steps 4 through 7 several times to verify that a leak is indic-
ated. If a leak is indicated, call a Micromeritics Service Representative.
8. Press 9 to close the Vent valve, 8 to open the Expansion valve, and 7 to open the Fill valve;
O - O - X.
9. Fill the chambers to 19.5 psig.
10. Press 7 to close the Fill valve; X - O - X.
11. Observe the pressure display. After an equilibration period (about 20 to 30 seconds), the pres-
sure should not decrease more than 0.007 psig (0.048 kPag) or increase more than 0.02 psig
(0.14 kPag) in a two minute span.
If the pressure decreases more than 0.007 psig (0.048 kPag) or increases more than 0.02
psig (0.14 kPag) in a two minute span, temperature instability or a leak may be indicated. Vent
the system, then repeat steps 8 through 11 several times to verify that a leak is indicated. If a
leak is indicated, call a Micromeritics Service Representative.
12 - 8
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals