12 Maintenance and Troubleshooting
SET REGULATOR PRESSURE WITH THE KEYPAD
The Purge fill pressure and cycle fill pressure are specified using Setup > Analysis Parameters. Set
the regulator pressure at the higher of the two (plus 2.0 psig). For example; if the purge fill pressure
is specified as 19.500 psig and the cycle fill pressure as 19.000 psig, then set the regulator pressure
at 21.500 psig.
1. Press Alt + 1 to enter Manual mode.
2. Press 8 (EXPAND) and 9 (VENT) to open the expansion and vent valves; the display will
show: X - O - O
3. Press 7 (FILL) to open the fill valve; the display will show: O - O - O
4. Adjust the regulator pressure control knob until the Purge fill or Cycle fill pressure, whichever
is higher, is shown on the regulator display.
5. Press 7 to close the Fill valve, then increase the regulator valve by 2.0 psig. The display will
show: X - O - O.
6. Allow the pressure in the pycnometer to drop below 2.0 psig, then press 8 and 9 to close the
Expansion and Vent valves. The display will show: X - X - X
7. Press SAVE to return to the Reload prompt.
12 - 16
AccuPyc II 1345 Operator Manual
134-42851-01 (Rev A) Jan 2021

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micromeritics ACCUPYC II 1345 and is the answer not in the manual?

Micromeritics ACCUPYC II 1345 Specifications

General IconGeneral
BrandMicromeritics
ModelACCUPYC II 1345
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals