LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
Section Menu Section Contents Procedures Back Index
317
8
14. Disconnect the battery pins/power PCB cable (W10) (3207692-000) from the rear case by removing the four Kep nuts
(201508-000) (see Figure 8.99 on p. 305). Reuse the Kep nuts.
15. Remove the CO2 exhaust (3012140-002) from the rear case (if installed).
16. Remove the USB flex assembly as described in Removing the W14 - USB Flex Module (p. 339)
~ Remove the drain seal (3006291-00) from the rear case.
17. Remove the handle as described in Removing the Handle (p. 321).
To remove the rear case: (Continued) 17 steps, (Page 2 of 2)