LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
Section Menu Section Contents Procedures Back Index
316
8
Removing the Rear Case
To remove the rear case: 17 steps, (Page 1 of 2)
1. Disassemble the case as described in Disassembling the Case (p. 181).
2. Remove the system/therapy PCB assembly as described in System (A01)/Therapy (A04) PCB Assembly Replacement (p. 232).
3. Remove the screw (202253-761) securing the clear, plastic, high-voltage shield (3010593-00) from the interconnect bracket
(A17) (see Figure 8.2 on p. 178). Discard the screw.
4. Remove the two mounting screws (202253-761) from the capacitor bracket and remove the bracket (3207031-001) (see
Figure 8.65 on p. 269). Discard the screws.
5. Remove the two biphasic module mounting screws (3207337-312) (see Figure 8.95 on p. 300). Discard the screws.
6. Observing orientation, cut the large tie wrap securing the A15 energy storage capacitor and interconnect bracket to the rear
case. Remove the energy storage capacitor, interconnect bracket, transfer relay and Biphasic module assembly without
removing small cable ties and spade terminal connections.
7. Remove the OEM PCB as described in Removing the OEM PCB (A06) (p. 262).
8. Remove the parameter bezel as described in Removing the Parameter Bezel (p. 311).
9. Remove the NIBP/CO2 module as described in Removing the NIBP (A21)/CO2 (A23) Modules (p. 288). (Removing the bracket
cover in this procedure is not required.)
10. Remove the SpO2 module as described in Removing the SpO2 PCB (A16) (p. 275) (if installed).
11. Remove the power PCB as described in Removing the Power PCB (A03) (p. 254).
12. Remove the contact PCB as described in Removing the Contact PCB (A07) (p. 359).
13. Remove the system connector cable and auxiliary connector cable as described in Removing the System Connector Cable (W08)
or Auxiliary Power Cable (W09) (p. 332) (use new O-ring seals (200060-011) during installation).