LIFEPAK 15 Monitor/Defibrillator
Service Manual
Replacement Procedures
Rear Case Replacement
Section Menu Section Contents Procedures Back Index
318
8
Installing the Rear Case
To install the rear case: 23 steps, (Page 1 of 3)
• When replacing the Rear Case, use Rear Case Repair Kit (MIN 3305431-024) (p. 512).
• Clean adhesive areas with alcohol.
Transfer the following parts from the old rear case to the new rear case:
1. Install handle (3207706-000), handle covers and shoulder straps as described in Handle Replacement (p. 321).
NOTE: If a new handle is required, use Handle Repair Kit (MIN 3305431-020) (p. 509).
2. Install the drain seal (3006291-00) into the rear case.
3. Install the CO2 exhaust tubing (3012140-002) onto rear case (if CO2 option).
4. Install the contact PCB as described in Installing the Contact PCB (A07) (p. 359).
5. Install the system connector and auxiliary connector cables as described in Installing the System Connector Cable (W08) or
Auxiliary Power Cable (W09) (p. 333) using new O-ring seals (200060-011).
6. Install the EMI shield (3206960-002) as described in EMI Shield Replacement (p. 305) (new part from repair kit).
7. Install the battery pins/power PCB cable (W10) (3207692-000) as described in Installing the Battery Pins/Power PCB Cable
(W10) (p. 336).
8. Install the power to contact PCB cable (W05) (3207261-001) to J42 of the contact PCB.
9. Install the USB flex assembly as described in USB Flex Module (W14) Replacement (p. 338).