Chapter E: CALIBRATIONS 2 CALIBRATING DYNAMIC EXPOSURE CONTROL (DEC)
Technical manual Planmeca ProOne 113
Enter DEC calibration display as follows:
Touch the service spanner icon on the main display. The main display is the display that is
shown when the X-ray unit is switched on. Select the option Calibrations (T3400) on the
Technical setting
s display. On the next display, select the option DEC calibration (T3440).
The
DEC calibration
(T3440)
display will be shown.
NOTE
You need a password to gain access to the technical settings. The password needs to be
entered each time the X-ray unit is switched on. The password is 1701.
NOTE
DEC calibration will be shown on the
Calibrations (T3400)
display only if DEC has been
activated on the
Program licences (P2100)
display.
Service spanner