4 TECHNICAL SETTINGS Chapter B: SETTINGS
72 Planmeca ProOne Technical manual
• mA R´10 series in use
This option allows you to introduce the mA R´10 series instead of the normal mA
values (2, 3, 4, 5, 6, 7). The mA values in R´10 series are: 2.2, 2.8, 3.5, 4.4, 5.6, and
7.0. These values will appear on the kV and mA display accessed through the kV/mA
field on the main display.
Touch the
mA R´10 series in use
field in order to activate this option.
Changing the mA scale will automatically restore the factory settings for the quick button
selections made by user through the Main display / kV/mA field / parameter table / quick
button settings at the bottom of the parameter table.
• Moving column (T3710)
The moving column moves at two different paces called the
initial speed
(slower, not
adjustable), and
final speed
(faster). This option allows you to adjust he maximum
pace of the
final speed
separately for the upwards and downwards movement as
described below.
The initial speed (slower) will not be affected by this selection.
Touch the right/left arrow button next to
Max speed up
to adjust the maximum speed
of upwards movement within range 50-100% in steps of 5%.
Touch the right/left arrow button next to
Max speed down
to adjust the maximum
speed of downwards movement within range 10-100% in steps of 5%.
You can test the changed speed with the up/down keys on the display. The speed of
upwards/downwards movement will change accordingly.
Accept your selection by touching the green check mark button in the bottom right
corner of the display. To exit the display without saving the changes, touch the red
cross button in the bottom left corner of the display.
test upwards
test downwards
movement
movement